Installation of Lakeshore 370 AC Bridge with the Dilution Refrigerator

Hardware:

[1] Connect NI GPIG-ENET/1000 to the Lakeshore 370 with a GPIB cable.

[2] Connect the NI GPIG-ENET/1000 to the Instrument Ethernet switch 
      (switch IP range: 192.168.10.X).

[3] Plug in both devices and turn them both on.

[4] The NI GPIG-ENET/1000 is ready for communication when the Ready light stops flashing and becomes steady-on. The ACT/LINK light should also be on if the connection is complete. (The NI GPIG-ENET/1000 may be pinged at IP: 192.168.10.xxx (get IP address written on the ENET/1000 to verify connection to Instrument PC.)

[5] NOTE: The ENET/1000 must be set up in NI-MAX and MAX will assign a GPIB address that must agree with the visa_resource_name in the controller.config file.		           Example: visa_resource_name = GPIB3::13::INSTR 



Software: 

Enabling the SPICE drivers:
In controllers.config	find  [lakeshore_370_ac_bridge] and set enable = true

In motors.config	find  [lakeshore_370_setpoint]   and set enable = true

In parameters.config	find  [lakeshore_370_sensor]      and set enable = true



Alias/Command to set the temperature setpoint:
set_tempdr <value in Kelvin>

Example: To set the temperature setpoint to 30 mK: 
set_tempdr .03



SPICE Indicators:
The temperature setpoint will show in motors as: dr_setpoint
The current temperature will show in motors and parameters as: dr_tsample



NOTE:  Config file templates and alias may be found in the Add into  files at:
C:\SPICE\dev\spice_code\common\devices\lakeshore_370_ac_bridge\